LCR-6000 Series User Manual
103
FREQuency Subsystem
The FREQuency command sets the oscillator frequency. The FREQuency?
query returns the current test frequency setting.
Figure 11-4 FREQ Subsystem Command Tree
Command
Syntax
FREQ[:CW] {<value>,MIN,MAX}
Parameter
Where,
<value> Is the numeric data (NR1 integer, NR2fix float
or NR3 floating point).
MIN Sets to the minimum value
MAX Sets to the maximum value
Example
SEND> FREQ 1K<NL>//Set to 1kHz,the Hz cannot
be added.
Query Syntax
FREQ[:CW]?
Query Response
<NR3><NL>
NR3 floating point
Example
SEND> FREQ?<NL>
RET> 1.000000E+03<NL>
Note
A suffix multiplier (k) can be used with this command.
But the suffix unit Hz cant be used.
This command CANNOT be used in LIST SWEEP
DISPLAY page and CORRECTION page.
FREQuency
[:CW]
{value,MIN,MAX}