11.10.6 FETCh:LIST?
The FETCh:LIST? query sets the latest LIST measurement data of the
primary parameters, secondary parameters and comparator results into the
output buffer. Only applicable when in the [LIST MEAS] page view.
Query Syntax
FETCh:LIST?
Query Response
<spot no>,<NR3:primary value>,<NR3:secondary
value>,<CMP result> // spot no:01-10
Example
SEND> FETC:LIST?<NL>
RET> 01,-2.98524e-
12,+3.27673e+00,L,02,+7.11030e-12,+3.48450e-
01,P,03,+7.11322e-12,+5.14944e-02,H,04,-
1.00000e+20,-1.00000e+20,-,05,-1.00000e+20,-
1.00000e+20,-,06,-1.00000e+20,-1.00000e+20,-,07,-
1.00000e+20,-1.00000e+20,-,08,-1.00000e+20,-
1.00000e+20,-,09,-1.00000e+20,-1.00000e+20,-,10,-
1.00000e+20,-1.00000e+20,-<NL>
(-1.00000e+20: The STEP is OFF)
Query Syntax
FETCh:LIST? <spot no>
Query Response
<spot no>,<NR3:primary value>,<NR3:secondary
value>,<CMP result>
Example
SEND> FETC:LIST? 2<NL>
RET> 02,+7.11030e-12,+3.48450e-01,P<NL>
(-1.00000e+20: The STEP is OFF)
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for GW Instek LCR Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek LCR Series and is the answer not in the manual?

GW Instek LCR Series Specifications

General IconGeneral
BrandGW Instek
ModelLCR Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals