EasyManua.ls Logo

GW Instek LCR Series - Fetch:list

GW Instek LCR Series
143 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
110
11.10.6 FETCh:LIST?
The FETCh:LIST? query sets the latest LIST measurement data of the
primary parameters, secondary parameters and comparator results into the
output buffer. Only applicable when in the [LIST MEAS] page view.
Query Syntax
FETCh:LIST?
Query Response
<spot no>,<NR3:primary value>,<NR3:secondary
value>,<CMP result> // spot no:01-10
Example
SEND> FETC:LIST?<NL>
RET> 01,-2.98524e-
12,+3.27673e+00,L,02,+7.11030e-12,+3.48450e-
01,P,03,+7.11322e-12,+5.14944e-02,H,04,-
1.00000e+20,-1.00000e+20,-,05,-1.00000e+20,-
1.00000e+20,-,06,-1.00000e+20,-1.00000e+20,-,07,-
1.00000e+20,-1.00000e+20,-,08,-1.00000e+20,-
1.00000e+20,-,09,-1.00000e+20,-1.00000e+20,-,10,-
1.00000e+20,-1.00000e+20,-<NL>
(-1.00000e+20: The STEP is OFF)
Query Syntax
FETCh:LIST? <spot no>
Query Response
<spot no>,<NR3:primary value>,<NR3:secondary
value>,<CMP result>
Example
SEND> FETC:LIST? 2<NL>
RET> 02,+7.11030e-12,+3.48450e-01,P<NL>
(-1.00000e+20: The STEP is OFF)
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for GW Instek LCR Series

Related product manuals