Trigger mode 
When entering LIST MEAS for the first time, the system will 
set Trig Mode to MAN and automatically performs a 
measurement. 
Internal Trigger. All ten list points 
are swept continuous. 
Manual Trigger. Each time the 
instrument is triggered by [Trig] 
key, the list points are swept one by 
one. 
External Trigger. Each time the 
instrument is triggered by the 
handler trigger pin, the list points are 
swept one by one. 
BUS Trigger. Each time the 
instrument is triggered by SCPI 
command, the list points are swept 
one by one. 
The [LIST MEAS] page will execute a list of up to 10 
swept frequency or amplitude tests. 
When [MODE] is set to SEQ and [TRIG] is set to 
MAN, the [LIST MEAS] function will automatically 
execute each test step on the list in sequence until the 
last step is executed. The meter will then wait for the 
trigger button to be depressed before repeating the 
process. 
When [MODE] is set to STEP and [TRIG] is set to 
MAN, the [LIST MEAS] function will automatically 
execute the first test step. The meter will then wait for 
the trigger button to be depressed before executing the 
next step. This process is repeated for each step on the 
list. 
Sequence measure all step by one 
trigger. 
Measure one step by one trigger. 