Command Reference
120
Query Response
{on,off}
Example
SEND> CORR:SPOT:STATe?<NL>
RET> on<NL>
11.13.10 CORRection:SPOT:FREQuency
The CORRection:SPOT:FREQuency command sets the frequency point for
the specified frequency point correction.
Command
Syntax
CORRection:SPOT:FREQuendy <value>
Parameter
value NR1,NR2 or NR3:Frequecny value.
A suffix multiplier can be used with this command. But
the unit Hz” cannot be added.
Example
SEND> CORR:SPOT:FREQ 100<NL>
SEND> CORR:SPOT:FREQ 10k<NL>
Query Syntax
CORRection:SPOT:FREQuency?
Query Response
<NR3>
Example
SEND> CORR:SPOT:FREQ?<NL>
RET> 1.000000e+03<NL>
11.13.11 CORRection:SPOT:OPEN
This command executes the OPEN correction data measure for the specified
frequency correction.
Command
Syntax
CORRection:SPOT:OPEN
Example
SEND> CORR:SPOT:OPEN<NL>
RET> pass/fail<NL> //Finish or Undone
11.13.12 CORRection:SPOT:SHORt
This command executes the SHORT correction data measure for the
specified frequency correction.
Command
Syntax
CORRection:SPOT:SHORt
Example
SEND> CORR:SPOT:SHOR<NL>
RET> pass/fail<NL> //Finish or Undone