Setup Key
58
5.2.1 Measurement Function [FUNC]
The LCR-6000 Series simultaneously measures four
components of the complex impedance (parameters) in a
measurement cycle. These include primary parameter,
secondary parameter and two monitor parameters. Please
refer to the 4.1 [MEAS DISPLAY] Page section on 24 for
details.
5.2.2 Comparator Function ON/OFF
LCR-6000 Series built-in comparator can sort DUTs
into a maximum of 10 bins (BIN1 to BIN9 and OUT),
using a maximum of nine pairs of primary parameter
limits and one pair of secondary parameter limits.
A DUT can be sorted into the auxiliary BIN (AUX)
when the primary parameter is within the limits but the
secondary parameter is not.