LCR-6000 Series User Manual
117
11.13 CORRection Subsystem
The CORRection subsystem command group sets the correction function,
including the OPEN, SHORT and SPOT correction settings.
Note
The CORRection subsystem CANNOT work in
[LIST MEAS] page.
Figure 11-10 CORRection Subsystem Command Tree
11.13.1 CORRection:OPEN
The CORRection:OPEN command execute all preset OPEN correction data
measurement points.
Command
Syntax
CORRection:OPEN
Example
SEND> CORRection:OPEN<NL>
RET> open<NL> // Run
CORRection
:
OPEN
:
STATe
{ON,OFF or 1,0}
null
:
LCR
null
:
DCR
null
:
SHORt
:
STATe
{ON,OFF or 1,0}
null
:
LCR
null
:
DCR
null
:
SPOT
:
STATe
{ON,OFF or 1,0}
:
OPEN
null
:
SHORt
null
FREQuency:
<value>