LCR-6000 Series User Manual 
105 
11.8.2 LEVel:CURRent (=CURRent[:LEVel]) 
The LEVel:CURRent or CURRent[:LEVel] command sets the oscillator’s 
output current level.  
LEVel:CURRent {<value>,MIN,MAX} 
or CURRent:LEVel {<value>,MIN,MAX} 
Where,  
<value>   Is the numeric data (NR1, NR2 or NR3).  
MIN  Sets to the minimum value of current. 
MAX Sets to the maximum value of current. 
SEND>  LEV:CURR 1m<NL>//Set to 1mA. The A 
unit suffix cannot be used. 
LEVel:CURRent?  
or CURRent:LEVel? 
SEND>   CURR?<NL> 
RET>  1.000e+00<NL> 
The A unit suffix cannot be used. 
This command CANNOT be used in LIST MEAS page 
and CORRECTION page. 
11.8.3 LEVel:SRESistance (= VOLTage:SRESistance) 
The LEVel:SRESistance or VOLTage:SRESistance command sets the source 
output impedance. 
LEVel:SRESistance {30,50,100} 
VOLTage:SRESistance {30,50,100} 
{30,50,100} 
Where, 
30  Sets the output impedance to 30 
50  Sets the output impedance to 50 
100  Sets the output impedance to 100 
SEND>  LEV:SRES 30<NL>//Set to 30, the unit  
cannot be added. 
VOLTage:SRES? 
or LEVel:SRES? 
SEND>   LEV:SRES?<NL> 
RET>  30<NL> 