LCR-6000 Series User Manual
119
11.13.6 CORRection:SHORt:STATe
The CORRection:SHORt:STATe command sets the SHORT correction
function to ON or OFF.
CORRection:SHORt:STATe {ON,OFF or 1,0}
ON, 1 Set up the function is ON
OFF,0 Set up the function is OFF
SEND> CORR:SHOR:STATe ON<NL>
RET> null
SEND> CORR:SHOR:STATe?<NL>
RET> on<NL>
11.13.7 CORRection:SHORt:LCR
The CORRection:SHORt:LCR command execute all preset SHORT
correction data measurement points (AC).
SEND> CORRection:SHORt:LCR<NL>
RET> LCR short<NL> // Run
pass/fail<NL> //Finish or Undone
11.13.8 CORRection:SHORt:DCR
The CORRection:SHORt:DCR command execute all preset SHORT
correction data measurement points (DC).
SEND> CORRection:SHORt:DCR<NL>
RET> DCR short<NL> // Run
pass/fail<NL> //Finish or Undone
11.13.9 CORRection:SPOT:STATe
The CORRection:SPOT:STATe command sets the SOPT correction function
to ON or OFF.
CORRection:SOPT:STATe {ON,OFF or 1,0}
ON, 1 Set up the function is ON
OFF,0 Set up the function is OFF
SEND> CORR:SPOT:STATe ON<NL>
RET> null