EasyManua.ls Logo

GW Instek LCR Series - Level Subsystem; Level:voltage (=Voltage[:Level]); Figure ‎ 1 1-5 Level Subsystem Command Tree

GW Instek LCR Series
143 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
104
11.8 LEVel Subsystem
The Level subsystem sets the oscillators output voltage/current level and
source output impedance.
Figure 11-5 LEVel Subsystem Command Tree
11.8.1 LEVel:VOLTage (=VOLTage[:LEVel])
The LEVel:VOLTage or VOLTage[:LEVel] command sets the oscillators
output voltage level.
Command
Syntax
LEVel:VOLTage {<value>,MIN,MAX}
or VOLTage:LEVel {<value>,MIN,MAX}
Parameter
Where,
<value> Is the numeric data (NR1, NR2 or NR3).
MIN Sets to the minimum value of voltage.
MAX Sets to the maximum value of voltage.
Example
SEND> LEV:VOLT 0.3<NL>//Set to 0.3V, V unit
suffix cannot be used.
Query Syntax
LEVel:VOLTage?
or VOLTage:LEVel?
Query Response
<NR3>
NR3 floating point
Example
SEND> VOLT?<NL>
RET> 1.000e+00<NL>
Note
The V unit suffix cant be used.
This command CANNOT be used in LIST MEAS page
and CORRECTION page.
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for GW Instek LCR Series

Related product manuals