Command Reference
102
Query Response
<0-8><NL>
Example
SEND> FUNC:DCR:RANG?<NL>
RET> 0<NL>
11.7.5 FUNCtion:RANGe:AUTO
The FUNCtion:RANGe:AUTO command sets the auto range to ON or OFF.
Command
Syntax
FUNC:RANGe:AUTO {off(hold),on(auto),NOMinal}
Parameter
Where, {off(hold),on(auto),NOMinal} is:
off(or hold): Sets the auto range to off.
on(or auto): Sets the auto range to on.
NOMinal: Sets the range No.
Example
SEND> FUNC:RANG:AUTO AUTO<NL> //Sets to
auto range.
SEND> FUNC:RANG:AUTO off<NL> //Sets auto
range to off.
Query Syntax
FUNC:RANGe:AUTO?
Query Response
{HOLD,AUTO,NOM}
Example
SEND> FUNC:RANG:AUTO?<NL>
RET> auto<NL>
11.7.6 FUNCtion:MONitor 1 /2
The FUNCtion:MONitor1 and FUNCtion:MONitor2 commands set the two
monitor parameter.
Command
Syntax
FUNC:MONitor1 {off, Z, D, Q, THR, THD, R, X, G, B,
Y, ABS, PER VAC, IAC}
FUNC:MONitor2 {off, Z, D, Q, THR, THD, R, X, G, B,
Y, ABS, PER VAC, IAC}
Parameter
Where, {off, Z, D, Q, THR, THD, R, X, G, B, Y, ABS,
PER VAC, IAC}
Example
SEND> FUNC:MON1 Z<NL>
Query Syntax
FUNC:MON1?
FUNC:MON2?
Query Response
{off, Z, D, Q, THR, THD, R, X, G, B, Y, ABS, PER
VAC, IAC}
Example
SEND> FUNC:MON1?<NL>
RET> off<NL>
All manuals and user guides at all-guides.com
all-guides.com

Table of Contents

Other manuals for GW Instek LCR Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek LCR Series and is the answer not in the manual?

GW Instek LCR Series Specifications

General IconGeneral
BrandGW Instek
ModelLCR Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals