LCR-6000 Series User Manual
59
Figure 5-3 Page Comparator Workflow
Sorting Start
COMP=ON
BIN1MIN<Primary<BIN1MAX
.
BIN9MIN<Primary<BIN9MAX
AUX=ON
AUX=ON
End
Lower<Secondary<Upper
Display: OUT NG
Output: NG
P_OVER
P_HI or P_LO
No
No
No
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
Lower<Secondary<Upper
Display: OUT AUX NG
Output: NG
S_OVER
P_OVER
P_HI or P_LO
Display: BIN1-9 AUX NG
Output: NG
S_OVER
Display: BIN1-9 OK
Output: BIN_1 or BIN_2
Procedure for setting the comparator function
[COMP]
Step 1. Press the [Setup] key.
Step 2. Press the [BIN SETUP] soft-key.
Step 3. Use the cursor key to select [COMP] field.
Step 4. Use the soft-keys to turn on/off the
comparator.
Soft-key
Function
OFF
Turns OFF the COMP function
ON
Turns ON the COMP function