Measure Key
44
4.3.4 List Point and Limit Modes
The List measurement feature supports up to 10 list
points as well as measurement limit values. Each list
point can be turned on or off.
To configure the list points:
Step 1. Press the [Measure] key.
Step 2. Press the [LIST SETUP] soft-key.
Step 3. Use the cursor key to select one of the list
point field (1 through 10)
Step 4. Use the soft-keys to turn on/off the current
point.
Soft-key
Function
ON
Turn on current point
OFF
Turn off current point
Step 5. Enter the list point value (maybe the
frequency value or the signal level
voltage/current value). When you enter the
frequency value, the soft-key labels change to
unit labels (Hz, kHz). When you enter the
signal level voltage value, you don’t need to
enter units. When you enter the signal level
current value, the soft-key labels change to
unit labels (uA, mA).
Step 6. Using the cursor keys, select the LMT field.
Step 7. Configure the limit parameter by pressing the
appropriate soft-key.
Soft-key
Function
PRIMARY A
Uses the primary parameter as the
limit parameter.
2NDARY B
Uses the secondary parameter as the
limit parameter.
OFF
Turns OFF the limit feature.
Step 8. Using the cursor keys, select the LOWER
field.
Step 9. Enter the lower limit value.
Step 10. Using the cursor keys, select the UPPER field
and enter the higher limit value.
Step 11. Repeat Step 4 to Step 10.