EasyManua.ls Logo

GW Instek LCR Series - Fetch Subsystem; Fetch

GW Instek LCR Series
143 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
108
11.10 FETCh Subsystem
The FETCh subsystem command group is a sensor-only command which
retrieves the measurement data taken by measurement(s) initiated by a trigger,
and places the data into the output buffer.
Figure 11-7 FETCh Subsystem Command Tree
11.10.1 FETCh?
The FETCh? query sets the latest measurement data of the primary, secondary
parameters and comparator result into the output buffer.
Query Syntax
FETCh?
Query Response
<NR3:primary value>,<NR3:secondary
value>,<comparator result>
Example
SEND> FETC?<NL>
RET> +2.61788e-11,+5.45442e-01,BIN1,AUX-
OK,OK <NL>
RET> +1.23434e+05,OUT ,NG <NL>//DCR &
Comp on.
When in the [LIST MEAS] page view, the following description for the
FETCh query is applicable:
The FETCh? query sets the latest LIST measurement data of the primary
parameters, secondary parameters and comparator results into the output
buffer.
Query Syntax
FETCh?
Query Response
<NR3:primary value>,<NR3:secondary value>,
<CMP result>
Example
SEND> FETC?<NL>
RET> -2.98524e-12,+3.27673e+00,L <NL>
RET> -1.00000e+20,-1.00000e+20,- <NL>//The
STEP is OFF
FETCh
MAIN?
MONitor1?
IMPedance?
MONitor2?