LCR-6000 Series User Manual
127
12. SPECIFICATION
Accuracy is defined as meeting all of the following
conditions:
Temperature:
23 ºC ±5 ºC
Humidity:
<70% R.H.
Zeroing:
Open and Short Correction Warm
up time is 30 min or more.
1-year calibration cycle.
Basic Accuracy:
0.05%(Slow/Med), 0.1%(Fast)
12.1 General Specification
Display:
RGB color TFT-LCD, Size: 3.5”
(320x240)
Test Function:
Cs-Rs, Cs-D, Cp-Rp, Cp-D, Lp-
Rp, Lp-Q, Ls-Rs, Ls-Q,
Rs-Q, Rp-Q, R-X, DCR, Z-θr, Z-
θd, Z-D, Z-Q
Monitor Parameter:
Z, D, Q, Vac, Iac, Δ, Δ%, θr, θd,
R, X, G, B, Y (2 Parameter)
Measurement Speed:
40 times/s, 10 times/s, 3 times/s