LCR-6000 Series User Manual
47
There are 4 available trigger modes in the [LIST MEAS]
page.
Trigger Modes
Function
INT
Trigger by the meter’s internal trigger
source.
MAN
Trigger by pressing the trigger button.
EXT
Trigger by sending a signal to the
trigger pin on the handler interface.
BUS
Trigger by the RS-232 port.
4.4.3 Range Mode [RANGE]
Displays the selected measurement range for each
measurement step; this field is here only for display, its
content cant be changed here. If you need to change the
measurement range; it can be done in the
[MEAS DISPLAY] or [MEAS SETUP] function.
4.4.4 Measurement log [LOG]
The LCR-6000 has an internal data buffer that record up
to 10000 measurement readings. These readings can be
saved to an external USB drive in a .csv file format.
These readings can then be opened on a PC using
software such as Windows Excel. When the LOG
function is active, the display is fixed to the [MEAS
DISPLAY] screen. The [FUNC], [FREQ] and [LEVEL]
settings are also fixed and cannot be changed.
Procedure for setting Log
Step 1. Press the [LIST MEAS] soft-key.
Step 2. Use the cursor keys to select the [LOG] status
field.
Step 3. Use the [START LOG] soft-keys to start a
new log.
Soft-key
Function
START LOG
To start a new measurement log.
Step 4. After a new log has started, the measurement
readings will be recorded into the meter’s
internal buffer. The options in the following
table are available during this recording
process: