Setup Key
60
5.2.3 Auxiliary Bin [AUX]
After AUX is turned on, DUTs that do not fall within
the primary parameter limit values are sorted as OUT. In
addition, DUTs that fall within the primary parameter
limits but are out of the secondary parameter limits are
sorted into the auxiliary (AUX) bin.
Procedure for turning ON/OFF the auxiliary bin [AUX]
Step 1. Press the [Setup] key.
Step 2. Press the [BIN SETUP] soft-key.
Step 3. Use the cursor key to select [AUX] field.
Step 4. Use the soft-keys to turn on/off the auxiliary
bin.
Soft-key
Function
ON
Turns ON the auxiliary bin
OFF
Turns OFF the auxiliary bin
5.2.4 Primary Parameter Comparator limit mode
[MODE]
You can specify the primary parameter limit values in
one of the following 3 ways:
Tolerance mode [%][Δ]
In tolerance mode, the comparison limit values are
based on the deviations from the specified nominal
value, which can be specified in the [NOM] field.
You configure the tolerance mode limit values as a
deviation percentage (%) or absolute (Δ) parameter
value.
Sequential mode [SEQ]
In sequential mode, the comparison limit values are
based on the absolute value of the measurement.
When you configure these limit values, you have to
first define the minimum value and then the
maximum value.
Nominal Value [NOM]
Sets the nominal value (reference value) for the ABS
and PER compare modes. Use the number keys to
enter the value and the unit.