Setup Key 
60 
5.2.3 Auxiliary Bin [AUX] 
After AUX is turned on, DUTs that do not fall within 
the primary parameter limit values are sorted as OUT. In 
addition, DUTs that fall within the primary parameter 
limits but are out of the secondary parameter limits are 
sorted into the auxiliary (AUX) bin. 
Procedure for turning ON/OFF the auxiliary bin [AUX] 
Step 1.  Press the [Setup] key. 
Step 2.  Press the [BIN SETUP] soft-key. 
Step 3.  Use the cursor key to select [AUX] field. 
Step 4.  Use the soft-keys to turn on/off the auxiliary 
bin. 
Turns ON the auxiliary bin 
Turns OFF the auxiliary bin 
5.2.4  Primary Parameter Comparator limit mode 
[MODE] 
 
You can specify the primary parameter limit values in 
one of the following 3 ways: 
  Tolerance mode [%][Δ]  
In tolerance mode, the comparison limit values are 
based on the deviations from the specified nominal 
value, which can be specified in the [NOM] field. 
You configure the tolerance mode limit values as a 
deviation percentage (%) or absolute (Δ) parameter 
value. 
  Sequential mode [SEQ]  
In sequential mode, the comparison limit values are 
based on the absolute value of the measurement. 
When you configure these limit values, you have to 
first define the minimum value and then the 
maximum value. 
  Nominal Value [NOM] 
Sets the nominal value (reference value) for the ABS 
and PER compare modes. Use the number keys to 
enter the value and the unit. 