Query Response
<function>
Example
SEND> FUNC?<NL>
RET> Cp-D<NL>
11.7.2 FUNCtion:IMPedance:AUTO
The FUNCtion:IMPedance:AUTO command sets the impedances LCZ
Automatic selection.
Command
Syntax
FUNC:IMPedance:AUTO {ON,OFF, 0,1}
Example
SEND> FUNC:IMP:AUTO ON<NL>
Query Syntax
FUNC:IMPedance:AUTO?
Query Response
{on,off}<NL>
Example
SEND> FUNC:IMP:AUTO?<NL>
RET> off<NL>
11.7.3 FUNCtion:IMPedance:RANGe
The FUNCtion:IMPedance:RANGe command sets the impedance’s
measurement range.
Command
Syntax
FUNC:IMPedance:RANGe <0-8,MIN,MAX>
Parameter
Where, <0-8,MIN, MAX> is:
0-8 The range number
MIN =Range 0
MAX =Range 8
Example
SEND> FUNC:IMP:RANG 2<NL>//Set
measurement range to [2] 10k.
Query Syntax
FUNC:IMPedance:RANGe?
Query Response
<0-8><NL>
Example
SEND> FUNC:IMP:RANG?<NL>
RET> 0<NL>
11.7.4 FUNCtion:DCR:RANGe
The FUNCtion:DCR:RANGe command sets the DCRs measurement range.
Command
Syntax
FUNC:DCR:RANGe <0-8,MIN,MAX>
Parameter
Where, <0-8,MIN, MAX> is:
0-8 The range number
MIN =Range 0
MAX =Range 8
Example
SEND> FUNC:DCR:RANG 2<NL>//Set DCR range
to [2] 10k.
Query Syntax
FUNC:DCR:RANGe?
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for GW Instek LCR Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek LCR Series and is the answer not in the manual?

GW Instek LCR Series Specifications

General IconGeneral
BrandGW Instek
ModelLCR Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals