5.1.2 Averaging Factor [AVG]
The averaging feature allows you to obtain moving
average values of successive measurement results.
You can specify the averaging factor from integer 1 to
integer 256.
To set up the averaging factor
Step 1. Press the [Setup] key.
Step 2. Use the cursor key to select [AVG] field.
Step 3. Use the soft-keys or number keys to enter
averaging factor.
Soft-key
Function
INCR +
Increments the averaging factor in
steps of 1, 2, 4, 8, 16, 32,
64, 128 and 256.
DECR -
Decrements the averaging factor in
steps of 1, 2, 4, 8, 16, 32,
64, 128 and 256.
5.1.3 DC Bias Voltage [BIAS]
This function will apply a DC bias voltage onto the DUT
while LCR meter is performing its AC measurement on
the DUT. The settable DC bias voltage range is from -
2.5V ~ 2.5V.
Procedure for setting DC bias voltage
Step 1. Press the [Setup] key.
Step 2. Use the cursor key to select [BIAS] field.
Step 3. Use the soft-keys to select DC bias voltage.
Soft-key
Function
OFF
Turn off the DC bias voltage
2V
2V voltage is applied to the device under
test.
1.5V
1.5V voltage is applied to the device under
test.
-1.5V
-1.5V voltage is applied to the device under