LCR-6000 Series User Manual 
55 
  Procedure for setting monitor parameters (monitor 1 
and monitor 2) 
Step 1.  Press the [Setup] key. 
Step 2.  Use the cursor key to select [MON 1] or 
[MON 2] field. 
Step 3.  Use the soft-keys to select DC bias voltage. 
Absolute value of impedance 
Absolute value of admittance 
5.1.6 Measurement Delay [DELAY] 
This setting decides how long the meter will wait to 
commence a measurement after the trigger signal is 
received. Settable range: 0ms to 60s. 
5.1.7 AUTO LEVEL CONTROL [ALC] 
The automatic level control (ALC) feature adjusts the 
voltage across the DUT or the current through the DUT 
to match the voltage/current level setting. Using this 
feature, you can try to ensure a constant signal level 
(voltage or current) is applied to the DUT. 