Affinity® Four Birthing Bed Service Manual (195826 REV 2) 2 - 1
2
Chapter 2
Troubleshooting
GETTING STARTED
Only facility-authorized persons should troubleshoot the bed. Troubleshooting by unauthorized
persons could cause personal injury or equipment damage.
Begin each procedure in this chapter with step 1. Follow the sequence outlined (each step assumes the
previous step has been completed). In each step, the normal operation of the product can be confirmed
by answering Yes or No to the statement. Your response will lead to another step in the procedure, a
repair analysis procedure (RAP), or a component replacement. If more than one component is listed,
replace them in the given order.
To begin gathering information about the problem, start with Initial Actions.
To isolate or identify a problem and to make sure that a repair was done correctly after completing each
corrective action (replacing or adjusting a part, seating a connector, etc.), do the Function Checks.
To make sure that a repair was done correctly, do the Final Actions after the Function Checks.
If troubleshooting procedures do not isolate the problem, contact Hill-Rom Technical Support.