Function Checks
Chapter 2: Troubleshooting
Affinity® Four Birthing Bed Service Manual (195826 REV 2) 2 - 5
2
FUNCTION CHECKS
Function checks determine whether the bed is operating correctly. All caregiver control panel functions
are available from both the right and left siderails. When checking redundant function controls, activate
each of the siderails to determine if the fault is contained in one or both of the siderails.
1. Initial Actions have been done.
Yes No
Go to “Initial Actions” on page 2-4.
2. The bed is plugged into an appropriate power source.
Yes No
Plug the bed into an appropriate power source. Go to Step 3.
3. The lockout switch is in the “ON” (unlocked) position before proceeding with the function checks.
Yes No
Put the lockout switch in the “ON” (unlocked) position. Go to Step 4.
4. The optional pendant, if equipped, is connected correctly.
Yes No
Connect the pendant to the pendant connector, located on the bottom motor
cover. Go to Step 5.
5. Examine the siderail cable connections to the logic control P.C. board cable. The cables are
connected correctly.
Yes No
Connect the siderail cable to the cable from the logic control P.C. board. Go to Step
6.
6. Visually examine the bed for loose connections on cable assemblies and electrical wires. All cable
connections and wiring are correctly secured.
Yes No
Attach the cable connections and wiring. Go to Step 7.
7. Examine the bed for signs of obvious damage. The bed appears to be all right.
Yes No
Repair the damage, and go to “Hilow System Functional Check” on page 2-5.
HILOW SYSTEM FUNCTIONAL CHECK
1. Use the bed function controls to put the bed in these positions:
Put the bed in the mid-height position.
Raise the foot section to the high position.
Lower the head to the low position.
Make sure that the bed is out of the Trend-Like position (the sleep surface is level).
2. Momentarily press the Bed Up control. The bed rises.
Yes No
Go to RAP 2.4 on page 2-13.