4.18 Air System Manifold Assembly
Chapter 4: Procedures
4 - 32 Affinity® Four Birthing Bed Service Manual (195826 REV 2)
4.18 Air System Manifold Assembly
Tools: Torx® T25 driver
REMOVAL
1. Remove the air system assembly (see Procedure 4.17 on page 4-29).
2. Disconnect the air hoses from the manifold assembly (A) (see Figure 4-21 on page 4-32).
3. Disconnect the cable (B) from the air manifold to the logic control P.C. board connector P12.
4. Remove the two screws (C) that attach the manifold assembly (A) to the mounting plate (D).
Figure 4-21. Air System Manifold Assembly
REPLACEMENT
1. Do the removal procedure in reverse order.
2. Do the “Function Checks” on page 2-5.
4.19 Line Voltage Fuses
Tools: Small screwdriver
Two mains fuses protect the high voltage circuit. The fuses (line voltage fuses) are located in the fuse
holder in the power entry module at the head end of the bed.
Follow all electrical safety precautions when servicing the bed's electrical system. Failure to do so could
cause personal injury or equipment damage.