EasyManua.ls Logo

Hill-Rom Affinity P3700E - Head Section Mattress Assembly; Labor Grips

Hill-Rom Affinity P3700E
202 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.15 Head Section Mattress Assembly
Chapter 4: Procedures
Affinity® Four Birthing Bed Service Manual (195826 REV 2) 4 - 27
4
4.15 Head Section Mattress Assembly
Tools: None
REMOVAL
1. Set the brake/steer pedal to the brake position.
2. Make sure the bed is out of the Trend-Like position.
3. Carefully lift the mattress from the head end of the bed.
CAUTION:
Do not pull on the mattress material when unfastening the mattress retaining snaps. Unfasten the
mattress retaining snaps at the snap location. Failure to do so could cause equipment damage.
4. Locate and unfasten the two mattress retaining snaps beneath the seat section.
5. If the mattress has air bladders, disconnect the two hoses located beneath the seat section.
REPLACEMENT
1. Do the removal procedure in reverse order.
2. Make sure the two air hoses are correctly installed below the seat section.
3. Do the “Function Checks” on page 2-5.
4.16 Labor Grips
Tools: T25 Torx® head bit Ratchet Lubri-film spray (P/N 36958)
External snap-ring removal tool Plug insertion tool (P/N 151085)
REMOVAL
1. Set the brake/steer pedal to the brake position.
2. Make sure the bed is out of the Trend-Like position.
3. Remove the mattress to gain access to the seat panel. See Procedure 4.15 on page 4-27.
4. Remove the screws (A) that attach the seat panel (B) to the seat pan (C) (see Figure 4-17 on page 4-
28).
5. Remove the plug (G).