Operating tabletop mechanical components
4990851_002_00 – 2082262 – 2022-01-10 33
3. Mount the edging assembly on top of the tabletop (4 screws).
Tighten the screws with a torque of 10 Nm.
4. Remove the cable ties from the edging assembly and column.
5. Install the retaining plate W401 (2 screws).
6. Attention! Do not interchange the plug connections on the
circuit board. Pay attention to the wiring diagram on page 64.
Restore the original cable routing. Connect the plug of cable
W402 to the PCBA distributor board and mount the cable
shield.
7. Attention: Do not interchange the plug connections. Pay
attention to the wiring diagram on page 64.
Restore the original cable routing. Insert the plug of cables
W201, W301 and W305 into the W401 cable bushing.
8. Mount protective conductor W404 on the column head (1
screw with 1 washer and 2 serrated lock washers).
9. Mount the column head cover (Chapter 9.3).
10. Mount the pad plate onto the strut (Chapter 7.2).
11. Guide the column cover upwards and close (Chapter 6.6).
12. Put the pad in place.
13. Connect the power supply to the operating table
(Chapter 6.2).