Column electric components
64
4990851_002_00 – 2082262 – 2022-01-10
4. If all work on the lift is complete, fasten the cover plate [3] to
the spindle end (securing ring pliers and omniFIT 100M thread
locking agent).
5. Attach the underbody cover on the running gear (7 screws).
6. Glue a new protective sheet over the bearing housing of the
spindle.
7. Restore the operating table to the upright position
(Chapter 6.8).
8. Put the pad in place.
9. Connect the power supply to the operating table
(Chapter 6.2).
10. Update the firmware of the operating table.