Column electric components
4990851_002_00 – 2082262 – 2022-01-10 71
5. Mount the 2 cable clips above the Main Controller Unit (1
screw each).
6. Reattach any cable fastenings that have been removed.
7. Guide the column cover upwards and mount it at the head
and foot end of both guide column brackets (2 screws each).
8. Return the flat conductor to its original mounting position.
Pull the bellows upwards and push the flat conductor of the
column keypad into the plug of the main board (Main
Controller Unit) and engage. Push the small pins at the edge
of the plug upwards for this purpose.
9. Mount the lower frame of the bellows on both guide column
brackets at the head and foot end (2 screws each).
10. Mount the bellows on the column cover (1 screw each on the
left and right below the bellows).
11. Put the pad in place.
12. Connect the power supply to the operating table
(Chapter 6.2).
13. Configure the new main electronic unit with the service
software.
14. Update the firmware of the operating table.