11.2 Control and Monitoring Using a LAN Interface
173
Connection to the instrument
3. Confirm the IP address and subne
t
mask for the computer you are using
.
Do not check the [Obtain an IP addres
s
automatically] option.
To set the instrument’s IP address
,
change the final numerical value only fo
r
the computer’s IP address.
To set the instrument’s subnet mask, se
t
the same value as the computers subne
t
mask.
Example:
PC
IP address: 192.168.0.2
Subnet mask: 255.255.255.0
3196
IP address: 192.168.0.1
Subnet mask: 255.255.255.0
Start up Internet Explorer, and enter the IP
address that you set for the instrument (suc
h
as http://192.168.0.1) in the address column
,
then press the Enter key.
When a normal connection is established, th
e
Main page for the instrument’s HTTP serve
r
function opens.
Configuration of Internet Explorer:
Deselect the [Use proxy server] option.
Configure Internet Explorer so that it does no
t
use a proxy server by clicking [Interne
t
Options] on Internet Explorer’s [Tools] menu
,
then clicking the [Connections] tab.
If connection fails:
Check the following items.
Verify that you are using a straight LAN cable. If you are using a straight LAN cable, the com-
puter’s LINK LED lights.
Verify that you set the IP address and subnet mask.
Verify that the instrument’s RS connection point is the modem.
Verify that you are receiving a normal response from the ping command on the computer’s DOS
prompt screen.
Verify that the checkmark is removed from the Internet browser proxy setting.