12.1 Product Specifications
179
The specifications below apply to the 3196 POWER QUALITY ANALYZER.
For the product specifications of the EN50160, see the Instruction Manual for EN50160.
Specifications Chapter 12
12.1 Product Specifications
Environmental and Safety Specifications
Operating environment
Indoors, altitude up to 2000 m (6562-ft.)
Storage temperature and
humidity
-20°C to 50°C (-4°F to 122°F), 80% RH or less (no condensation)
Operating temperature and
humidity
0°C to 40°C (32°F to 104°F), 80% RH or less (non-condensating)
In the temperature range -10°C to 0°C(14°F to 32°F), battery pack and PC
card use are not supported. The LCD and accuracy are also not guaran-
teed.
Maximum input voltage
(50/60 Hz)
Voltage inputs: 780 Vrms, 1103 V peak value
Current inputs: 1.7 Vrms, 2.4 V peak value
Maximum rated voltage to
earth
Voltage input terminals: 600 Vrms(50/60 Hz)
Dielectric strength
(50 Hz for 1 min.)
5.55 kVrms for one minute (current sensitivity 1 mA)
Between voltage and clamp input terminals, between voltage input termi-
nals and instrument case, between voltage input terminals U1 to U3 and U4
Enclosure protection
IP30 (per EN60529)
Applicable Standards
Safety EN61010
Voltage Input: Pollution Degree 2,
Measurement Category III (anticipated transient overvoltage
6000V)
EMC EN61326 Class A
EN61000-3-2
EN61000-3-3