Appendix
213
Measurement Range and Nominal Voltage
The voltage and current ranges of this unit are as follows.
This unit is not equipped with an automatic range selection function,
therefore you must select the operation ranges.
The display and effective measurement ranges (ranges where accu-
racy is certain) of measurement ranges are as follows.
Voltage range
Appendix
Voltage
range
CH 1 to 3 150 V, 300 V, 600 V
CH 4 (AC) 60 V, 150 V, 300 V, 600 V
CH 4 (DC) 60 V, 600 V
Current
range
0.1 mV/A 500 A, 5000 A
1 mV/A (9661) 50 A, 500 A
1 mV/A (9660) 50 A, 100 A
10 mV/A (9694) 5 A, 50 A
100 mV/A 0.5 A, 5 A
5000 A 9667 500A, 5000A
500 A 9667 50A, 500A
1000 A 9669 100A, 1000A
有効測定範囲
600V
660V
780V
6V0.9V
346600V
0.1V 1.0V 10V 100V 1000V
600V
入力電圧
○電流レンジ
330V
有効測定範
300V
390V
3V0.45V
200277V
300Vレン
165V
有効測定範囲
150V
195V
1.5V0.23V
100120V
150V
有効測定範囲
60V
66V
78V
0.6V0.09V
60V
表示範囲
有効測定範囲
5000A
5500A
6500A
50A7.5A
5000Aンジ
Effective measurement range
Effective measurement range
Effective measurement range
Effective measurement range
-
-
-
Input Voltage
600 V
300 V
150 V
60 V
Display range
Set the nominal voltage so that it does not exceed the selected voltage range
.
You can use a nominal voltage of 346 to 600 V in a 600 V voltage range.
You can use a nominal voltage of 200 to 277 V in a 300 V voltage range.
You can use a nominal voltage of 100 to 120 V in a 150 V voltage range.

Table of Contents