5.6 Waveform Calculation
______________________________________________________________________________________________
5.6.2 Setting Type of Calculation
Setting Screen: Calc
CH1 (CH2)
NOTE
Sets the type of waveform calculation.
Setting Procedure
1. Press the
COMP SETTINGS
key, and
select "
Calc
"tag.
2. Select "
CH1
" on the menu, then press the
ENTER
key.
3. Use the
CURSOR
key to select an item,
then press the
ENTER
key. The
waveforms obtained by the calculations
below will be displayed on CH1.
OFF : Does not perform calculation.
ABS : Calculates Absolute value for
CH1.
Rectify : Calculates Half-wave
rectification for CH1.
CH1+CH2 : Calculates CH1+CH2.
CH1-CH2 : Calculates CH1-CH2.
CH1xCH2 : Calculates CH1xCH2.
CH1/CH2 : Calculates CH1/CH2.
4. When using the 8731-10, also be sure to
set the type for "
CH2
". The waveforms
obtained by the following calculations will
be displayed on CH2.
OFF : Does not perform calculation.
ABS : Calculates Absolute value for
CH2.
Rectify : Calculates Half-wave
rectification for CH2.
CH2+CH1 : Calculates CH2+CH1.
CH2-CH1 : Calculates CH2-CH1.
CH2xCH1 : Calculates CH2xCH1.
CH2/CH1 : Calculates CH2/CH1.
For the 8730-10, only "
ABS
" and "
Rectify
" are available.