31
_____________________________________________________________________________________________
3.2 Basic Measurement and Setting
______________________________________________________________________________________________
3.2.2 Measurement Examples
Configure
Time/DIV 200 ms
Shot 10 DIV
Time Zoom X1
Format Single
Average OFF
Overlay OFF
Info (Left) Gauge (CH1)
Info (Upper) OFF
Auto Save OFF
Unit (CH1)
Range 500 mV
Coupling DC
Filter 5kHz
Zoom X1
ZeroPos. +50%
Wave Red
Scaling OFF
Variable OFF
Trigger
Trigger Mode Repeat
Pre-Trigger 0%
Trigger Source OR
CH1 Level, 0 V, UP
CH2 OFF
External OFF
Timer OFF
Example 1) Connect an oscillator to the 8730-10, 8731-10 and measure 4 Vp-p
1 kHz sine waves.
1. Connect an oscillator to the 8730-10, 8731-10.
2. Set the oscillator to generate 4Vp-p 1kHz sine waves.
3. Set the measurement parameters.
4. Set the input parameters.
5. Set the trigger parameters.