8.5 Output Terminals
______________________________________________________________________________________________
8.5.1 Output of Value Comparison Results
<OK_VALUE Terminal (B2), NG_VALUE Terminal (B1)>
T T
BUSY
STORAGE
TRIG_OUT
JUDGE_OUT
OK_VALUE
NG_VALUE
Start
Operation o
the 8730-10
8731-10
Sto
Wait fo
tri
er
Storing
waveform
Comparison
rocessin
Dis
la
rocessin
Wait fo
tri
er
Storing
aveform
Clears previous
com
arison result
Out
uts com
arison result
Clears comparison
result
Trigger activated
Trigger activated
Measurement start
Outputs the result of value comparison.
The NG_VALUE terminal (terminal No.: B1) outputs the total comparison
of value comparisons No.1 to No.4.
The OK_VALUE terminal (terminal No.: B2) outputs the total comparison
result of value comparisons No.1 to No.4.
The total comparison result is PASS when the results of value
comparisons No.1 to No.4 are all PASS. If any result is FAIL, the total
comparison result is FAIL.
Each value comparison result (No.1 to No.4) can be outputted separately.
(See 8.5.8)
Output Signal
The NG_VALUE terminal (B1) outputs a low-level signal (active LOW)
when any result of value comparisons No.1 to No.4 is FAIL.
The OK_VALUE terminal (B2) outputs a low-level signal (active LOW)
when the results of value comparisons No.1 to No.4 are all PASS.
<When using an internal power supply>
HIGH level voltage: Max.: 5.5 [V], Min.: 3.5 [V] (When there is no load)
LOW level voltage : Max.: 1.0 [V], Min: 0 [V]
<When using an external power supply>
External power supply voltage V
EXT
: Max.: 30 [V] DC, Min.: 5 [V] DC
HIGH level voltage : Max.: V
EXT
[V], Min.: V
EXT
-1.0 [V] (When there is no load)
LOW level voltage : Max.: 1.0 [V], Min.: 0 [V]
<Timing>
1. Trigger Mode: Repeat, Auto Save: OFF