54
_____________________________________________________________________________________________
4.2 Details of Measuremen t Setting
______________________________________________________________________________________________
4.2.6 Overlay Function
Setting Screen: Configure
Overlay (OFF, ON)
Overlaid display
NOTE
Overlay is performed without clearing the currently displayed waveform (if
trigger mode is "
Repeat
"or"
Auto
").
This allows comparison to the immediately preceding waveform.
Setting Procedure
1. Press the
DISPLAY
key to select the
"
Configure
" tag. Press the
ENTER
key to
display the menu.
2. Move the cursor to "
Overlay
", and press
the
ENTER
key.
3. Select "
ON
"or"
OFF
", then press the
ENTER
key to confirm the overlay
function setting.
When the trigger mode is set to "
Single
",
overlay is not available since measurement
will end after loading data once.
The following changes are only reflected in
the waveform loaded last.
Changes in magnification or compression
rate (time axis or voltage axis)