52
_____________________________________________________________________________________________
4.2 Details of Measuremen t Setting
______________________________________________________________________________________________
4.2.5 Averaging
Setting Screen: Configure
Average (OFF, 2, 4, 8,16)
Waveform before averaging
Waveform after averaging
NOTE
The currently selected averaging count is shown in red.
If the trigger is not set properly, the intended result may not be obtained. (See 4.4)
For the calculation equations for averaging, see 13.2.4.
When avera
in
is enabled
waveform calculation
see 5.6
is disabled.
The averaging function allows capturing several instances of a waveform
and determining the average.
This makes it possible to eliminate noise and irregular signal components.
Setting Procedure
1. Press the
DISPLAY
key to select the
"
Configure
" tag. Press the
ENTER
key
to display the menu.
2. Move the cursor to "
Average
", and
press the
ENTER
key.
3. Select an averaging count, then press
the
ENTER
key to confirm.