8.5 Output Terminals
______________________________________________________________________________________________
8.5.6 Processing <BUSY Terminal <(B10)>
T T
TRIG_OUT
Operation o
f
the 8730-10
8731-10
Sto
p
Measurement start
Wait fo
r
tri
gg
er
Storing
w
aveform
Comparison
p
rocessin
g
Dis
p
la
yp
rocessin
g
Wait fo
r
tri
gg
er
Storing
w
aveform
Trigger
activated
Trigger
activated
T
BUSY
Operation o
f
the 8730-10
8731-10
Sto
p
Measurement start
Wait fo
r
tri
gg
er
Storing
w
aveform
Comparison
p
rocessin
g
Dis
p
la
yp
rocessin
g
Trigger
activated
Measurement sto
p
Sto
p
T T
BUSY
Operation o
f
the 8730-10
8731-10
Measurement start
Wait fo
r
tri
gg
er
Storing
w
aveform
Comparison
p
rocessin
g
Display
p
rocessin
g
Trigger
activated
Sto
p
Trigger
activated
Save
p
rocessin
g
Wait fo
r
tri
gg
er
Storing
w
aveform
T
BUSY
Operation o
f
the 8730-10
8731-10
Measurement start
Wait fo
r
tri
gg
er
Storing
w
aveform
Comparison
p
rocessin
g
Display
p
rocessin
g
Sto
p
Save
p
rocessin
g
Measurement sto
p
Sto
p
Trigger
activated
Indicates that the 8730-10, 8731-10 is conducting measurement and
processing a task (e.g., storing a waveform, comparison, saving data).
The BUSY terminal (terminal No.: B10) outputs a low-level signal (active
LOW) during 8730-10, 8731-10 operation or processing when it cannot
accept the START signal to start measurement.
Output Signal
The BUSY terminal (B10) outputs a low-level signal (active LOW) during
8730-10, 8731-10 processing.
<When using an internal power supply>
HIGH level voltage: Max.: 5.5 [V], Min.: 3.5 [V] (When there is no load)
LOW level voltage : Max.: 1.0 [V], Min: 0 [V]
<When using an external power supply>
External power supply voltage V
EXT
: Max.: 30 [V] DC, Min.: 5 [V] DC
HIGH level voltage : Max.: V
EXT
[V], Min.: V
EXT
-1.0 [V] (When there is no load)
LOW level voltage : Max.: 1.0 [V], Min.: 0 [V]
<Timing>
1. Trigger Mode: Repeat, Auto Save: OFF
2. Trigger Mode: Single, Auto Save: OFF
3. Trigger Mode: Repeat, Auto Save: ON
4. Trigger Mode: Single, Auto Save: ON