215
_____________________________________________________________________________________________
8.5 Output Terminals
______________________________________________________________________________________________
8.5.7 Measuring <STORAGE Terminal (B11)>
T
STORAGE
START
BUSY
Measurement start
Trigger activated
Measurement sto
Operation of
the 8730-10
8731-10
Sto
Wait for tri
e
Storin
waveform
Comparison
rocessin
Display
rocessin
Sto
Less than 85 ms
Less than 350 ms
Screen u
date
ait for amplifier signal/
tri
er
Storin
waveform
Storing waveform is possible
(Triggering is acceeptable.)
Less than 435 ms
NOTE
The conditions for the above timing example are Single screen format, with no
LAN connection and no A/B cursor display.
Indicates that the 8730-10, 8731-10 is storing a waveform.
The STORAGE terminal (terminal No.: B11) begins to output a low-level
signal (active LOW) when the 8730-10, 8731-10 starts measurement and
updates the screen, and continues signal output until the 8730-10, 8731-10
completes waveform storing.
When STORAGE terminal begins to output LOW signal, triggering is
acceptable.
Output Signal
The STORAGE terminal (B11) outputs a low-level signal (active LOW)
while the 8730-10, 8731-10 is storing a waveform. The terminal is also
set to LOW level while the 8730-10, 8731-10 is waiting for a signal from
the amplifier or a trigger.
<When using an internal power supply>
HIGH level voltage: Max.: 5.5 [V], Min.: 3.5 [V] (When there is no load)
LOW level voltage : Max.: 1.0 [V], Min: 0 [V]
<When using an external power supply>
External power supply voltage V
EXT
: Max.: 30 [V] DC, Min.: 5 [V] DC
HIGH level voltage : Max.: V
EXT
[V], Min.: V
EXT
-1.0 [V] (When there is no load)
LOW level voltage : Max.: 1.0 [V], Min.: 0 [V]
<Timing>