77
_____________________________________________________________________________________________
4.4 Trigger Functions
______________________________________________________________________________________________
4.4.2 Trigger Mode
Setting Screen: Trigger
Trigger Mode (Single, Repeat, Auto)
The trigger mode determines the way triggering is used to control
operation of the unit.
When all trigger sources are OFF, a recording operation begins
immediately (free-run operation).
Setting Procedure
1. Press the
DISPLAY
key and use the right
and left
CURSOR
keys to select the
"
Trigger
"tag.
2. Select "
Trigger Mode
", then press the
ENTER
key.
(Note 1)
3. Select an item, then press the
ENTER
key.
Single : Trigger is registered only once. 8730, 8730-10 starts waveform recording when triggering
occurs and continues for preset recording length. Measurement then ends automatically.
Repeat: Trigger is registered continuously. Unit is in trigger standby condition when trigger
conditions are not met. Measurement ends when
STOP
key is pressed.
Auto : Trigger is registered continuously. If trigger sources are not met within 1 second, waveform
recording starts automatically and continues for preset recording length. Measurement ends
when
STOP
key is pressed.