99
_____________________________________________________________________________________________
5.2 Comparison Details
______________________________________________________________________________________________
5.2.2 Screen Updating
Setting Screen: Comp
Refresh Disp (OFF, Stop Mode, Always)
5.2.3 Switching Test Modes
Setting Screen: Comp
Select Mode (MODE01 to MODE16)
NOTE
The 8730-10, 8731-10 may be set to update the screen during comparison.
Normally set the unit to "
Always
" (default).
Select "
OFF
" to shorten the comparison cycle for reducing the tact time.
Without screen updating, the display processing time (see figure in 8.5.2)
is reduced, thus reducing the comparison cycle.
Not updating the screen reduces the comparison cycle by 100 ms for the
8730-10 and 50 ms for the 8731-10. However, this also reduces the
output period ("L" pulse) of comparison results (PASS or FAIL). The "L"
part of output is at least 10 ms. For details, see 8.5.2.
Setting Procedure
1. Press the
COMP SETTINGS
key, and
select "
Comp
"tag.
2. Use the
CURSOR
key to select "
Refresh
Disp" on the menu.
3. Use the
CURSOR
key to select a
condition, then press the
ENTER
key.
OFF : Does not update the screen.
Stop Mode : Updates the screen only
when the comparison result
satisfies the stop conditions.
Always : Always updates the screen.
Switches to the test mode saved in 8730-10, 8731-10 internal memory and
has been registered according to 5.2.4.
The registered test mode settings are read into the 8730-10, 8731-10.
When "
Info (Upper)
" is set to "
Comment
" (see 4.2.8), the comments
entered at mode registration appear in the file comment field.
Setting Procedure
1. Press the
COMP SETTINGS
key, select
"
Comp
" tag, then press the
ENTER
key.
2. Select "
Select Mode
" on the menu, and
press the
ENTER
key.
3. Use the
CURSOR
key to select a test
mode from "
MODE01
"to"
MODE16
"on
the menu, then press the
ENTER
key.
On the menu, the saved test modes are
shown in black. A mode must be saved
before it can be selected.