187
_____________________________________________________________________________________________
7.5 External Control Terminals
______________________________________________________________________________________________
Setting Screen: EXT-IN, OUT
EXT_OUT1 to EXT_OUT5 (Total, Area, Value)
When value comparison is selected,
select the type of calculation by number
Procedure
1. Press the
ADJUST/SYSTEM
key to
display the Adjustment screen.
2. Use the
CURSOR
key to select "
EXT-IN,
OUT
"on"
Page
". You can also use the
ADJUST/SYSTEM
key to change the
menu page. (Pressing this key moves to
next page.)
3. Use the
CURSOR
key to select
"
EXT_OUT1
", then press the
ENTER
key.
4. Use the
CURSOR
key to select an output
condition, then press the
ENTER
key.
Total : Result of all area and value
comparisons
Area : Result of area comparisons (CH1
x CH2)
Value : Result of individual value
comparison
5. Move the
CURSOR
key to "
PASS/FAIL
",
then press the
ENTER
key.
6. Use the
CURSOR
key to select "
PASS
"
or "
FAIL
", then press the
ENTER
key.
PASS : Outputs the signal when the result
set in 4. above is "
PASS
".
FAIL : Outputs the signal when the result
set in 4. above is "
FAIL
".
7. When value comparison is selected,
select the type of calculation by number.
Press the
ENTER
key. After selecting a
No., press the
ENTER
key once again.
8. When setting other terminals
(EXT_OUT2 to EXT_OUT5), repeat
steps 3. to 7. above.