81
_____________________________________________________________________________________________
4.4 Trigger Functions
______________________________________________________________________________________________
トリガレベル
フィルタ幅
ここではトリガがかかりません
ここでトリガが
かかります
T
Filter mar
in
Trigger Level
Triggering does not occur here.
A trigger is activated here.
NOTE
Increases in number
Move the cursor to right
Decreases in number
Editing Trigger Level
Move the cursor to left
Confirms the value
Range Setting range Step
100 mV -1 V to + 1.1 V 0.0025 V
200 mV -2 V to + 2.2 V 0.005 V
500 mV -5 V to + 5.5 V 0.0125 V
1V -10Vto+11V 0.025 V
2V -20Vto+22V 0.05 V
5V -50Vto+55V 0.125 V
(Example) The trigger slope is set to UP and the trigger filter is used.
Setting the Level Trigger
1. On the trigger condition menu of a
channel (CH1 or CH2), select "
Level
",
then press the
ENTER
key.
2. Set the trigger level under "
Level
".
The setting range of the trigger level varies
depending on the input signal range. The
value input for the level setting will also vary
depending on the range.