Revision 8 SLG 700 SmartLine Level Transmitter User’s Manual Page 87
Figure 3-41: Anchoring wire probes
When anchoring, a wire probe it is recommended that the wire be slack to prevent
excessive tensile loading from motion of medium and/or thermal expansion. The sag
should be ~1cm/m (1.5/10) of the probe length. See Figure 3-42.
Figure 3-42: Wire probe slack