Revision 8 SLG 700 SmartLine Level Transmitter User’s Manual Page 173
Ambient and Process Temperature Ranges
The ambient temperature range (Ta), maximum process temperature, and applicable temperature class
of the equipment is as follows:
SLG720
Protection Type
Ta, Ambient
Temperature Range
Maximum Process
Temperature
Temperature
Class
Intrinsic Safety
-50°C to +70°C
150°C
T4
Intrinsic Safety
-50°C to +60°C
200°C
T4
other protection types
-50°C to +85°C
150°C
T5
other protection types
-50°C to +70°C
150°C
T6
other protection types
-50°C to +60°C
200°C
T6
SLG726
Protection Type
Ta, Ambient
Temperature Range
Maximum Process
Temperature
Temperature
Class
Intrinsic Safety
-50°C to +70°C
300°C
T4
Intrinsic Safety
-50°C to +60°C
450°C
T4
other protection types
-50°C to +85°C
300°C
T5
other protection types
-50°C to +70°C
300°C
T6
other protection types
-50°C to +60°C
450°C
T6
All models when installed as FISCO Ta= -50°C to 45°C
Operating range is always limited to -40°C for all transmitters.
Models with CCoE, SAEx, KOSHA, NEPSI, INMETRO, or EAC TR-CU nameplate are limited to
temperature class T4 for all ambient and process temperature ranges. At the time of writing, only
CSA, FM, IECEx, and ATEX models have been certified with the temperature classes T5 or T6 as
specified in the table above.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell SmartLine SLG 700 and is the answer not in the manual?

Honeywell SmartLine SLG 700 Specifications

General IconGeneral
BrandHoneywell
ModelSmartLine SLG 700
CategoryTransmitter
LanguageEnglish

Related product manuals