8 Operation
98
Procedure for manual linking of digital sensors
Step Action
1 In the "Digital Sensors" menu, highlight the assignment table entry
to be linked.
2 Open the "Link" submenu by tapping the corresponding button in
the "Digital Sensors" menu.
3 Open the "bus scan" by pressing the "Scan" button (magnifying
glass).
4 In the list of detected sensors, highlight the one that you wish to link
with the previously highlighted assignment table entry. Sensors
with JUMO digiLine electronics can be identified by the hardware
address on the nameplate of the JUMO digiLine electronics. It is
also displayed in the list of detected sensors and in the sensor in-
formation ("Info" button).
5 The "Link" button now appears above the table. Press this button
to link the selected assignment table entry with the sensor selected
from the list of detected sensors.
6 If the link was successful, the previously selected entry now has
the status "linked" in the assignment table. If the digital sensor
could not be linked, an error message appears. The error message
displays a table with information about the selected assignment ta-
ble entry and the sensor selected from the list of detected sensors.
The error message is displayed in the lower area.
Example of an error
message:
No assignment table
entry was selected be-
fore the "Link" sub-
menu was opened.