19 Calibrating O-DO sensors
258
19.3 O-DO calibration routine
19.3.1 End value calibration
NOTE!
You must be logged in with corresponding user rights to perform calibrations.
Chapter 8.2.1 "Log-on/Log-out", page 89
NOTE!
A digital sensor can only be calibrated if it is linked.
Chapter 8.2.7 "Digital sensors", page 95
Step Action
1 Start the end value calibration.
Device Menu > Calibration > Digital Sensor 1 to 6 > End Value
Calibration
2 If temperature compensation was not specified in the calibration
default settings, enter the temperature of the air-saturated water or
the water vapor-saturated air here manually.
If temperature compensation was specified, the
temperature is determined automatically.
3 Immerse the sensor in air-saturated water or position it in water va-
por-saturated air (see Chapter 19.2.1 "Calibration methods for O-
DO sensors", page 256). If you use water vapor-saturated air, en-
sure constant air pressure, constant temperature and that no water
drops adhere to the membrane during the calibration procedure.
Wait until the measuring value displayed stabilizes and then con-
firm the result of the measurement by pressing "OK".
tap button to enter
temperature