EasyManua.ls Logo

JUMO AQUIS touch P - Calibration Methods for O-DO Sensors; Calibration Default Settings for O-DO Sensors

JUMO AQUIS touch P
404 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19 Calibrating O-DO sensors
256
19.2.1 Calibration methods for O-DO sensors
End value calibration
With end value calibration, the slope of the sensor is calibrated beyond the de-
fined state of 100 % oxygen saturation. This state can in principle be achieved
in two ways:
By positioning the sensor in water vapor-saturated air (for example, directly
over a water surface).
By positioning the sensor in air-saturated water (air is directed through water
until the water is saturated with it).
Two-point Calibration
With two-point calibration, the zero point and slope of the sensor are calibrated.
This calibration method offers the greatest possible level of accuracy and is par-
ticularly recommended for measurements of small oxygen concentrations.
19.2.2 Calibration default settings for O-DO sensors
Before you can perform a calibration, you must first enter the necessary calibra-
tion default settings. The possible settings for the O-DO calibration are de-
scribed in the following.
Open the calibration default settings:
Device Menu > Calibration > Digital Sensor 1 to 6 >
Calibration Default Settings
NOTE!
The "Calibration default settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in. Calibration default settings of a
digital sensor only appear for linked digital sensors.
Chapter 8.2.1 "Log-on/Log-out", page 89
Chapter 8.2.7 "Digital sensors", page 95
Sample screen:
O-DO calibration default
settings

Table of Contents

Related product manuals