10 Configuration
142
Connection type 2-wire,
3-wire
only for signal types Pt100, Pt1000, 400 Ω and
4000 Ω:
Connection variants for the connected resistance
thermometer
Customer-specific
linearization
Selection of a
linearization table
Linearization tables contain up to 40
value pairs in any measurement characteristic
line.
Each value pair assigns a display value
(Y-column) to a measuring value (X-column).
Up to 8 linearization tables can be stored. To cre-
ate a linearization table, you need the JUMO PC
setup program.
Unit up to 5 text characters Process variable unit
not adjustable for pH measurement
The temperature unit is specified in the basic set-
tings.
⇨ Chapter 10.2 "Basic settings", page 134
Scaling start -99999 to +99999
b
only for standard signals:
Measured value from sensor (uncomp.) that cor-
responds to the lower limit of the standard signal
range [0 V or 0(4) mA];
Refer to the technical data for
the sensor.
Scaling end -99999 to +99999
b
only for standard signals:
only for standard signals:
Measured value from sensor (uncomp.) that cor-
responds to the upper limit of the standard signal
range [10 V or 20 mA]
Refer to the technical data for
the sensor.
Start of
display range
-99999 to +99999
b
Upper/lower limit for labeling the scale when dis-
playing measured values, e.g. in recorder dia-
grams and bar graphs
Display range end -99999 to +99999
b
Decimal place Auto
Fixed comma format
Decimal points in the display
Offset -999 to +999
b
only for temperature and
conductivity measurement:
Correction value added to measured value
Filter Time Constant 0.0 to 25.0 s Optimization of measured value updating
The higher the value of the filter time const., the
slower the measured value is updated.
Ra 0 to 4000 Ω only for IN 11/12: Resistance value between the
slider (S) and start point (A) in a resistance po-
tentiometer when the slider is at the start point
Configuration point Selection/
setting option
Explanation