10 Configuration
152
Customer-specific
linearization
Selection of a
linearization table
Linearization tables contain up to 40
value pairs in any measurement characteristic
line.
Each value pair assigns a display value
(Y-column) to a measuring value (X-column).
Up to 8 linearization tables can be stored. To cre-
ate a linearization table, you need the JUMO PC
setup program.
Start of
display range
-99999 to +99999
a
Upper/lower limit for labeling the scale when dis-
playing measured values, e.g. in recorder dia-
grams and bar graphs
Display range end -99999 to +99999
a
Decimal place Auto
Fixed comma format
Decimal points in the display
Offset -99999 to +99999
a
Correction value added to measured value
Alarms 1/2
per measuring range
1to4
Analog input alarms are used to monitor measured values
in relation to adjustable limit values.
The alarm settings for all analog device functions are explained together.
⇨ Chapter 10.11.2 "Alarms for analog signals and digital sensors", page
170
a
The unit set for the conductivity measurement input appears in the entry field.
Configuration point Selection/
setting option
Explanation