8 Operation
96
Link status symbols for the assignment table entries
Status Explanation Symbol
linked Communication between the sensor electronics
and the JUMO AQUIS touch P has been suc-
cessfully established. The sensor is operating.
Installation The device searches for the individual assign-
ment table entry of a linkable sensor. A new sen-
sor can be linked manually or automatically if the
following prerequisites are met:
The sensor type must match the settings of
the assignment table entry.
The sensor version must be compatible
Operation of bus communication must be as-
sured through correct interface settings (with
JUMO digiLine electronics, the interfaces are
configured automatically during the scan)
Known sensors are identified and assigned and
linked with the known entry
After successful assignment, the status changes
from "Installation" to "linked" following connec-
tion of the sensor to the bus. If a sensor cannot
be linked, the status remains "Installation". If the
status has not changed from "Installation" to
"linked" after a waiting period of about 15 sec-
onds, linking of the sensor electronics has failed.
A description of the startup procedure
for digital sensors can be found in the chapter on
startup.
Chapter 7.2 "Digital sensors", page 68