Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-54 077085503 / March 2019
OUTPut subsystem
The output subsystem provides information and settings that control the source output.
:OUTPut:DELay:FALLing
This command specifies a delay that the instrument waits before turning off the specified output.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0 s
Usage
:OUTPut:DELay:FALLing <n>
:OUTPut:DELay:FALLing <n>,<channel>
:OUTPut:DELay:FALLing?
:OUTPut:DELay:FALLing? <channel>
:OUTPut:DELay:FALLing? DEFault
:OUTPut:DELay:FALLing? DEFault,<channel>
:OUTPut:DELay:FALLing? MINimum
:OUTPut:DELay:FALLing? MINimum,<channel>
:OUTPut:DELay:FALLing? MAXimum
:OUTPut:DELay:FALLing? MAXimum,<channel>
[n]
Delay rising time: 0 to 10 s with the resolution of 0.001 s.
<channel>
The channel number (default value is CH1):
ï‚§ Channel 1: CH1
Channel 2: CH2
Details
This command sets the falling delay so that output will not turn off until the delay has elapsed. This
command affects on-to-off transitions, including the operation:
:OUTPut:STATe ON
Example
:OUTP:DELay:FALLing 1
Set the falling delay time to 1 s.
Also see
:OUTPut:DELay:RISing (on page 7-55)
:OUTPut[:STATe] (on page 7-59)

Table of Contents

Other manuals for Keithley 2280S-32-6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2280S-32-6 and is the answer not in the manual?

Keithley 2280S-32-6 Specifications

General IconGeneral
BrandKeithley
Model2280S-32-6
CategoryPower Supply
LanguageEnglish

Related product manuals