Section 7: SCPI command reference Series 2280 Precision DC Power Supply Reference Manual
7-70 077085503 / March 2019
:SENSe[n]:<function>:REFerence
This command specifies the relative offset value for the measurement.
Recall settings
Instrument reset
Power cycle
Usage
:SENSe[n]:CONCurrent:REFerence <ref>
:SENSe[n]:CURRent:REFerence <ref>
:SENSe[n]:VOLTage:REFerence <ref>
:SENSe[n]:CONCurrent:REFerence?
:SENSe[n]:CURRent:REFerence?
:SENSe[n]:VOLTage:REFerence?
Channel number: For single-channel instruments, the value is 1
The relative offset value:
ï‚§ Concurrent and current: -6.1 to 6.1 A for Model 2280S-32-6, -3.3 to
3.3 A for Model 2280S-60-3
ï‚§ Voltage: -32 to 32 V for Model 2280S-32-6, -60 to 60 V for Model
2280S-60-3
Details
This command specifies the relative offset value that can be applied to new measurements. When
relative offset is enabled, all subsequent measured readings are offset by the value that is set for this
command.
You can set this value, or have the instrument acquire a value. If the instrument acquires the value,
read this setting to return the value that was measured internally.
For concurrent measurements, the offset value is applied only to current measurements.
Example
Set the relative offset for current measurements to 1.
Also see
:SENSe[n]:<function>:REFerence:ACQuire (on page 7-71)
:SENSe[n]:<function>:REFerence:STATe (on page 7-72)